Stone, Keeron J ORCID: 0000-0001-6572-7874, Fryer, Simon M ORCID: 0000-0003-0376-0104, Ryan, Terrence and Stoner, Lee (2015) The validity and reliability of continuous-wave near -infrared spectroscopy for the assessment of leg blood volume during an orthostatic challenge. In: The British Association of Sport and Exercise Sciences, December 2015, St Georges Park, UK. (Unpublished)
There is a more recent version of this item available. |
Text
Validity and reliability of continuous-wave.pdf Restricted to Repository staff only Download (468kB) |
Abstract
Near-infrared spectroscopy (NIRS) can potentially be used to assess the cardiovascular autonomic system by monitoring orthostatic challenge-induced shifts in lower limb blood volume. However, in order to be of clinical utility the test must be valid, reliable, and relatively simple to conduct. Purpose: To induce lower limb blood volume shifts using a 10min 70o head-up tilt, and: (1) in the soleus, determine the validity of an inexpensive continuous wave (cw)-NIRS device by comparing to a criterion frequency-domain (fd-) NIRS device, (2) determine the between-day reliability of soleus assessments obtained from cw-NIRS and fd-NIRS; and, (3) compare the between-day reliability for fd-NIRS assessments obtained at the soleus (standard) and gastrocnemius (simpler alternative). Methods: Fifteen non-smoking healthy adults were tested on 3 different mornings, under standardized conditions, separated by a maximum of 7 days. Total haemoglobin concentration (tHb) was continuously monitored bi-laterally in the medial soleus using cw-NIRS and fd-NIRS. For site comparison, tHb was measured in the medial gastrocnemius using fd-NIRS. Results: (1) The area under the curve (AUC) for cw-NIRS and fd-NIRS assessments at the soleus were not significantly different (p =.619). (2) The criterion (0.75) intra-class correlation coefficient (ICC) was exceeded for both cw-NIRS and fd-NIRS. (3) The criterion ICC was exceeded for both soleus and gastrocnemius assessments. Conclusion: Continuous-wave NIRS can be used to monitor orthostatic stress-induced shifts in lower leg blood volume with acceptable validity and reliability. This orthostatic test may present a relatively simple and inexpensive approach for assessing the cardiovascular autonomic nervous system.
Item Type: | Conference or Workshop Item (Paper) |
---|---|
Additional Information: | Article publication in Atherosclerosis see eprint.... |
Subjects: | G Geography. Anthropology. Recreation > GV Recreation Leisure > GV557 Sports R Medicine > RC Internal medicine > RC1200 Sports Medicine |
Divisions: | |
Research Priority Areas: | Health, Life Sciences, Sport and Wellbeing |
Depositing User: | Anne Pengelly |
Date Deposited: | 22 Jun 2016 09:54 |
Last Modified: | 27 Nov 2021 11:59 |
URI: | https://eprints.glos.ac.uk/id/eprint/3648 |
Available Versions of this Item
- The validity and reliability of continuous-wave near -infrared spectroscopy for the assessment of leg blood volume during an orthostatic challenge. (deposited 22 Jun 2016 09:54) [Currently Displayed]
University Staff: Request a correction | Repository Editors: Update this record